NIBP Sphygmomanometer Repeatability Tester & Calibrator

Jiangjing Technology IQS8808 Automated Non-Invasive Blood Pressure Monitor Repeatability Testing & Calibration Device

This device complies with YY9706.230-2023 Medical Electrical Equipment — Part 2-30: Particular Requirements for Basic Safety and Essential Performance of Automated Non-Invasive Blood Pressure Monitors, the pharmaceutical industry standard of the People's Republic of China.
As specified in clause 201.12.1.107:
The laboratory repeatability of blood pressure measurement for automated non-invasive blood pressure monitors shall be ≤ 3.0 mmHg (0.4 kPa).
Static pressure repeatability testing is performed 10 times at 250 mmHg, 150 mmHg and 50 mmHg.
Static pressure indication error verification points shall cover the entire measurement range with a minimum of 5 test points.


Product detail

Product Overview

The IQS8808 Automated Non-Invasive Blood Pressure Monitor Repeatability Testing & Calibration Deviceby Jiangjing Technology complies with YY9706.230-2023 Medical Electrical Equipment — Part 2-30: Particular Requirements for Basic Safety and Essential Performance of Automated Non-Invasive Blood Pressure Monitors, the pharmaceutical industry standard of the People's Republic of China.
As specified in clause 201.12.1.107:
The laboratory repeatability of blood pressure measurement for automated non-invasive blood pressure monitors shall be ≤ 3.0 mmHg (0.4 kPa).
Static pressure repeatability testing is performed 10 times at 250 mmHg, 150 mmHg and 50 mmHg.
Static pressure indication error verification points shall cover the entire measurement range with a minimum of 5 test points (e.g., 50 mmHg, 100 mmHg, 150 mmHg, 200 mmHg, 250 mmHg, 280 mmHg).
For the repeatability testing and calibration of automated non-invasive blood pressure monitors, Xiamen Jiangjing Technology Co., Ltd.has developed a device that provides a stable air source and simultaneously outputs 6 channels of different pressure values, enabling simultaneous repeatability testing and calibration of multiple non-invasive blood pressure monitors.
The blood pressure monitor repeatability testing & calibration bench has filed one invention patentand three utility model patents, filling a gap in the industry. It provides testing and calibration services for blood pressure monitor manufacturers, blood pressure monitor type approval authorities, blood pressure monitor verification institutions and other relevant organizations.

Key Features

  1. Equipped with high-precision pressure regulating valvesfor stable pressure output

  2. 6-channel independent pressure outputwith different pressure values

  3. Simple operationfor the complete system