As specified in clause
201.12.1.107:
The laboratory repeatability of blood pressure measurement for automated non-invasive blood pressure monitors shall be ≤
3.0 mmHg (0.4 kPa).
Static pressure repeatability testing is performed 10 times at 250 mmHg, 150 mmHg and 50 mmHg.
Static pressure indication error verification points shall cover the entire measurement range with a minimum of 5 test points (e.g., 50 mmHg, 100 mmHg, 150 mmHg, 200 mmHg, 250 mmHg, 280 mmHg).