Automatic NIBP Sphygmomanometer Tester / Calibrator

Product detail

Product Overview

The IQS1600 Fully Automatic Automated Non-Invasive Blood Pressure (NIBP) Monitor Verification Deviceis designed and manufactured in accordance with JJF 1626-2017 Calibration Specification for Blood Pressure Simulatorsand JJG 692-2010 Verification Regulation of Non-Invasive Automated Sphygmomanometers.
It is used for the verification of oscillometric non-invasive blood pressure monitors, providing dynamic blood pressure simulation, static calibration, automatic leak testing and over-pressure release testing. It can verify the performance indicators of various blood pressure monitors, and is suitable for blood pressure monitor calibration and testing in metrology institutes, hospitals, and electronic blood pressure monitor manufacturers' laboratories.

Product Features

  • Built-in automatic air pump

  • Built-in high-precision pressure sensor

  • 7-inch full touchscreen operation

  • Dynamic blood pressure (BP) simulation for arm-type and wrist-type cuff monitors

  • Systolic pressure simulation up to 280 mmHg

  • Compact size, light weight

  • Verifies non-invasive automated blood pressure monitors (including electronic blood pressure monitors and BP monitors), as well as conventional mercury sphygmomanometers and blood pressure gauges

  • Automatic measurement functions for pressure indication, air-tightness, deflation rate, sensitivity and pressure release

  • Two pressure control modes (pressurization and depressurization), with audible prompt when automatically adjusting to target pressure


Technical Parameters

  1. Range:(0 ~ 60) kPa (0 ~ 450 mmHg)

  2. Accuracy class:0.1% FS / 0.05% FS

  3. Pressure resolution:0.01 mmHg

  4. Absolute error:< 0.5 mmHg (0–300 mmHg)

  5. Pulse rate range:30 bpm – 250 bpm

  6. Pulse volume:0.10 cc – 1.20 cc

  7. Dynamic blood pressure simulation:

    • Adult mode:Systolic (50 ~ 255) mmHg ((6.7 ~ 34.0) kPa), Diastolic (30 ~ 195) mmHg ((4.0 ~ 26.0) kPa)

    • Neonatal mode:Systolic (30 ~ 120) mmHg ((4.0 ~ 16.0) kPa), Diastolic (10 ~ 100) mmHg ((1.3 ~ 13.3) kPa)

    • Blood pressure indication repeatability:≤ 0.13 kPa (1 mmHg)

    • Pulse rate range:(30 ~ 250) beats/min; Pulse rate error:±(1% reading + 1 digit)

    • Pulse volume:(0.1 ~ 2.4) mL; Pulse increment:0.1 mL

    • Pulse rate 200 beats/min at pulse volume 2.4 mL; Pulse rate 250 beats/min at pulse volume 1.2 mL

    • Leak rate:≤ 0.4 mmHg/min at 500 mL volume, 300 mmHg; ≤ 1 mmHg/min at 100 mL volume, 300 mmHg

  8. Air-tightness:≤ 0.3 kPa/min (2 mmHg/min)

  9. Dynamic BP simulation:8 standard adult BP simulations, 8 pathological BP models, and 6 neonatal BP simulations available

  10. 8 custom dynamic BP functions— user-defined systolic pressure, diastolic pressure and pulse rate parameters

  11. Built-in and external cuffoptions available

  12. Multi-function devicewith pressure gauge testing and pressure output functions

  13. Multiple pressure unitsincluding kPa, mmHg and psi

  14. Environmental conditions:

    • Ambient temperature: (20 ± 10) °C

    • Relative humidity: ≤ 80%

    • Power supply: (220 ± 22) V, (50 ± 1) Hz

  15. Dimensions / Weight:285 × 215 × 130 mm / 3.6 kg



Blood Pressure Monitor Verifier

In accordance with national metrological verification regulations JJG 270-2008and JJG 692-2010, this device uses microprocessor chips and pneumatic components to achieve intelligent detection, improving the detection accuracy of blood pressure monitors (gauges) and reducing labor intensity. It is suitable for testing mercury sphygmomanometers (gauges) and electronic blood pressure monitors (gauges), and is used by metrology departments, blood pressure monitor manufacturers and hospitals.

Product Features

  • Large LCD screen for easy operation

  • Automatic rapid inflation and pressure adjustment to required pressure points; fast and smooth pressure regulation with audible prompt

  • Safe detection even at the 40 kPa (300 mmHg) point for mercury sphygmomanometers

  • Sensitivity verification automatically processed per metrological verification regulations

  • Air-tightness verification with automatic calculation and display of results

  • Imported pressure sensor

  • Built-in overpressure protection (42 kPa) for safety and reliability

  • Quick selection of test points and test items; settings saved permanently after one configuration


Technical Parameters

  1. Measurement range:(0 ~ 60) kPa

  2. Pressurization range:(0 ~ 40) kPa

  3. Accuracy:Class 0.1

  4. Verification speed:(3–5) s per point

  5. Ambient temperature:(-5 ~ 50) °C

  6. Relative humidity:< 95%

  7. Dimensions / Weight:210 × 104 × 44 mm / 1 kg