Divided-flow humidity generator principle— shortens verification time and improves work efficiency
Software supportfor remote operation, scheduled start, remote power on/off, and fully automatic control — making calibration easier
Multiple protection functionsincluding over-temperature protection, automatic liquid replenishment, and dry-burn protection — ensuring safer calibration
Wide correction point rangewith smaller deviation
Automatic fluctuation calculationwith automatic stability detection and alert
Capacitive touchscreenwith one-touch screenshot
Automatic recordingof operational data and curve data
Real-time displayof temperature and humidity curves
| Item | IQSLH015 | IQSLH565 | IQSLH865 |
|---|---|---|---|
| Temperature Range | (5 ~ 55) °C | (-5 ~ 65) °C | (-30 ~ 80) °C |
| Temperature Uniformity | ≤ 0.3 °C (@ 15 °C, 20 °C, 30 °C) | ≤ 0.2 °C | |
| Temperature Fluctuation | ≤ ±0.1 °C / 30 min | ≤ ±0.05 °C | |
| Temperature Resolution | 0.01 °C | 0.01 °C | |
| Humidity Range | 10% RH ~ 95% RH (@ 20 °C) | 5% ~ 95% RH (@ 20 °C) | |
| Humidity Uniformity | ≤ 1.0% RH (@ 20 °C, 40% RH, 60% RH, 80% RH) | ≤ 0.5% RH | |
| Humidity Fluctuation | ≤ ±0.8% RH / 30 min (@ 20 °C, 40% RH, 60% RH, 80% RH) | ≤ ±0.3% RH | |
| Humidity Resolution | 0.01% RH | 0.01% RH | |
| Internal Working Dimensions | 510 × 510 × 490 / 800 × 800 × 490 mm | 500 × 500 × 600 mm | |
| External Dimensions | 1000 × 710 × 1800 / 1300 × 1000 × 1800 mm | 1050 × 1900 × 1800 mm | |
| Weight | 350 kg | 400 kg |
Dew point temperature sensor range:(-20 ~ 95) °C
Measurement accuracy:±0.2 °C
High resolution:0.01 °C
Temperature sensor range:(-40 ~ 60) °C
Accuracy:±0.1 °C
High resolution:0.01 °C
Repeatability:≤ ±0.05 °C
Cooling method:Two-stage semiconductor Peltier cooling
Real-time full-range curve display
Touchscreen operationwith Chinese-English dual interface, easy to operate
Automatic balancing systemfor short measurement time
Mirror contamination self-detectionwith manual mirror cleaning
Data storage, query and printing functions
Excellent long-term stability and repeatability, suitable for online analysis
Signal output:Serial communication RS232
Power supply:220 VAC, 50/60 Hz; Power:60 W
Operating environment:Temperature (-20 ~ 50) °C; Humidity ≤ 95%
Measurement range:(-20 ~ 100) °C
Measurement accuracy:≤ ±0.05 °C, high-precision platinum RTD sensor
Resolution:0.001 °C
Built-in 3000 mAh rechargeable batterywith standby time up to 70 hours
LCD display panelshowing real-time current temperature, battery level and resistance value
Display dimensions:100 mm (L) × 65 mm (W) × 32 mm (H)
Sensor dimensions:Length 460 mm, diameter 5 mm
Weight:270 g
Customizable soft-wire platinum RTD sensoravailable