Aneroid Barometer Calibration Tester
The IQS890 Aneroid Barometer Verification Solution is designed in accordance with JJG 272-2024 Verification Regulation for Aneroid Barometers and Aneroid Barographs.
The solution mainly consists of an automatic pressure calibrator, a 0.02-grade precision digital pressure gauge, an aneroid barometer verification chamber, and a high-low temperature test chamber. It is used for the verification and calibration of general aneroid barometers both on-site and in the laboratory.
According to measurement ranges, these instruments include conventional aneroid barometers (800 hPa ~ 1060 hPa), plateau aneroid barometers (500 hPa ~ 1060 hPa or 500 hPa ~ 1030 hPa), and aneroid barographs (870 hPa ~ 1050 hPa).


Product detail

Solution Overview

The IQS890 Aneroid Barometer Verification Solutionis designed in accordance with JJG 272-2024 Verification Regulation for Aneroid Barometers and Aneroid Barographs.
The solution mainly consists of an automatic pressure calibrator, a 0.02-grade precision digital pressure gauge, an aneroid barometer verification chamber, and a high-low temperature test chamber. It is suitable for the verification and calibration of general aneroid barometers both on-site and in the laboratory.
By measurement range, these instruments include:
  • Conventional aneroid barometers(800 hPa ~ 1060 hPa)

  • Plateau aneroid barometers(500 hPa ~ 1060 hPa or 500 hPa ~ 1030 hPa)

  • Aneroid barographs(870 hPa ~ 1050 hPa)


Solution Features

  • AC/DC poweredfor negative pressure (vacuum) generation

  • Built-in rechargeable lithium batterywith 16,800 mAh capacity — over one week of continuous operation on a single charge

  • CID control:Layered micro-touch switch — internal high pressure automatically triggers CID to activate pressurization, with explosion-proof protection

  • PTC control:In the event of overcurrent or short circuit, temperature rises sharply, PTC resistance increases, cutting off the current and stopping output

  • PME control:When temperature exceeds 160°C, PME micropores close, physically isolating and stopping current flow

  • Five-fold protectionon the current control board: overcurrent protection, short-circuit protection, low-voltage protection, overcharge protection, and temperature control

  • Automatic pressure generationwith a range of 500 hPa ~ 1200 hPa

  • External precision digital pressure gaugefor convenient calibration submission; pressure type: absolute pressure; range: (0 ~ 120) kPa.a; accuracy: 0.02%

  • Aneroid barometer verification chamber dimensions:160 × 125 mm (customizable)

  • Power source:Built-in electric pump with pressure regulation; power supply: DC 12V, 100W or AC 220V

  • Output ports:2 × M20×1.5 female-thread output ports; weight: 5 kg


Notes

I.In accordance with JJG 272-2024 Verification Regulation for Aneroid Barometers and Aneroid Barographs, a high-low temperature test chamber must be configured with temperature adjustment ranges at both ends: (0 ~ 5) °C and (25 ~ 35) °C; temperature fluctuation ≤ ±0.5°C; uniformity ≤ 1°C.
II.The pressure generation equipment can be replaced by a manual pneumatic pump / manual vacuum pump, configured with a 0.02% precision digital pressure gauge and an aneroid barometer verification chamber.