Clause 6.2.1 Measurement Standards:Standard thermometer (one of the measurement standards)
Clause 6.2.2 Other Equipment:Dry block calibrator or thermostatic bath (horizontal temperature difference ≤ 0.1 °C, maximum temperature difference ≤ 0.2 °C)
Clause 7.2.2 Calibration Point Selection — Temperature Indication Error:At least 4 calibration points, including -40 °C, 0 °C, 25 °C and 100 °C
Clause 7.2.18 Calibration Method for Temperature Indication Error:A multifunction temperature process calibrator shall be selected as the standard for calibrating temperature secondary instruments; a portable intelligent dry block calibrator shall be selected as the standard for calibrating temperature sensors.
Temperature range:(-40 ~ 100) °C
Specification requirement:Horizontal temperature difference ≤ 0.1 °C, maximum temperature difference ≤ 0.2 °C
Jiangjing Technology specification:Horizontal temperature difference ≤ 0.05 °C, maximum temperature difference ≤ 0.1 °C
Measurement range:(-40 ~ 100) °C
Measurement accuracy:≤ ±0.05 °C
High-precision platinum RTD sensor
Resolution:0.001 °C
Accuracy class:0.05%
Supports measurement and output of multiple parameters including current, voltage, resistance, frequency, thermocouple and RTD — a thermal secondary instrument calibration device